Real-time MRI-based control of a ferromagnetic core for endovascular navigation.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 18595804.
- Also identified by DOI 10.1109/TBME.2008.919720.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper shows that even a simple proportional-integral-derivative (PID) controller can be used in a clinical MRI system for real-time navigation of a ferromagnetic bead along a predefined trajectory. Although the PID controller has been validated in vivo in the artery of a living animal using a conventional clinical MRI platform, here the rectilinear navigation of a ferromagnetic bead is assessed experimentally along a two-dimensional (2D) path as well as the control of the bead in a pulsatile flow. The experimental results suggest the likelihood of controlling untethered microdevices or robots equipped with a ferromagnetic core inside complex pathways in the human body.
Medical subject headings
- Algorithms
- Blood Vessels
- Magnetic Resonance Imaging
- Magnetics
- Transducers