Median and ulnar nerve conduction studies at the wrist: criterion validity of the NC-stat automated device.

Armstrong, Theodore N; Dale, Ann M; Al-Lozi, Muhammad T; Franzblau, Alfred; Evanoff, Bradley A · J Occup Environ Med · 2008

case_control · Level III

Where this comes from

Abstract

To compare results obtained with the NC-stat--an automated nerve testing device--to traditional nerve conduction studies relevant to carpal tunnel syndrome screening. Thirty-three subjects recruited from patients referred for electrodiagnostic testing were studied. Measurements including the distal motor latency (DML), distal sensory latency (DSL), and median-ulnar latency difference (MUD) were obtained by the NC-stat and by standard nerve conduction studies. With modifications to the NC-stat's suggested reference ranges, sensitivity with respect to the traditional results ranged from 93.8% (sensory MUD) to 100% (median DML and DSL) and specificity ranged from 84.6% (motor MUD) to 94.1% (sensory MUD). Sensitivity was as high or higher and specificity was lower when using the manufacturer's suggested cutoffs. The NC-stat appears to be a convenient and sensitive method for detecting median nerve pathology at the wrist.