Modulation transfer function-based assessment of temporal resolution: validation for single- and dual-source CT.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 18632531.
- Also identified by DOI 10.1148/radiol.2482072173.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to determine a manufacturer-independent quality assurance measurement for temporal resolution with a three-dimensional cardiac motion robot; validation was with single-source (SS) and dual-source (DS) computed tomography (CT). Image acquisition was performed by using standard cardiac protocols. Image contrast-based modulation transfer function (MTF) was assessed as function of time. For motion frequency of 60 beats per minute, MTF slightly decreased by 14% and 6% for SS CT and DS CT, respectively. For higher frequencies, a stronger decrease of MTF (eg, by 50% [SS CT] and 18% [DS CT] at 120 beats per minute) was detected. Effect of manufacturer's adaptive bisegment algorithm for SS CT and corresponding resonance effects of rotation time and heart rate were quantified. The robot-based approach is a reproducible, objective way to assess temporal resolution; it allows practical measurement of temporal resolution and comparison of CT scanners and protocols.