Fixed-angle screws vs standard screws in acetabular prosthesis fixation: a cadaveric biomechanical study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 18639433.
- Also identified by DOI 10.1016/j.arth.2008.04.019.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Secure fixation of acetabular components in total hip arthroplasty can be challenging. The purpose of this study was to perform biomechanical analysis of cup fixation strength using fixed-angle vs standard screw fixation. Multihole, porous-backed acetabular prostheses were implanted in both acetabuli of 8 cadaveric pelves using standard press-fit techniques. Fixed-angle screws were used on the left side, and standard cancellous screws were used in the right. The use of fixed-angle screws enhanced acetabular fixation substantially under subfailure cyclic loading conditions and load-to-failure. The triradiate screw configuration increases the bending moment required to fail the specimens as well. Fixed-angle screws may be useful for achieving rigid fixation of acetabular prostheses in challenging clinical scenarios.
Medical subject headings
- Acetabulum
- Arthroplasty, Replacement, Hip
- Bone Screws
- Hip Prosthesis