Contact stress analysis of the conforming post-cam mechanism in posterior-stabilized total knee arthroplasty.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 18655950.
- Also identified by DOI 10.1016/j.arth.2007.05.023.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The present study evaluated the effects of extent of conformity of post-cam design on contact area and stress at post-cam mechanism using 4 different posterior-stabilized prostheses. TRAC and Alpina with full-conformed post-cams exhibited the largest contact area at 90 degrees and 120 degrees . PFC sigma RPF with partial conformed post-cam had the largest contact area at 150 degrees . Scorpio NRG with less conformed post-cam had smaller contact area than the others. Lifting of femoral component decreased contact area and increased contact stress of TRAC and Alpina. Recent modifications of post-cam design have increased contact area, contributing to lower contact stress. None of these prostheses exhibited constant low contact stress throughout flexion. Further modifications of post-cam mechanism are necessary to provide lower contact stress throughout deep knee flexion.