Effect of head surface roughness and sterilization on wear of UHMWPE acetabular cups.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 18671264.
- Also identified by DOI 10.1002/jbm.a.32161.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The impact of femoral head surface roughness on wear of gamma-irradiation sterilized (3 MRad in nitrogen, crosslinked) and nonsterilized (not crosslinked) UHMWPE acetabular cups has been evaluated. Gravimetric wear testing was performed in a hip joint simulator for 2 x 10(6) cycles. CoCrMo heads were used with different surface roughness (R(a) = 15 nm and R(a) = 400 nm). The surface roughness after wear test was unchanged for the roughened heads, whereas the initially smooth heads showed a few scratches. The roughened heads increased the wear of the acetabular cups 2-fold. The gamma-irradiated cups tested against rough heads underwent the highest wear. The absorption of water was highest for the gamma-irradiated cups (0.0204% compared to 0.0031% after 85 days). Raman spectroscopy showed small but significant crystallinity changes in the wear zone, where the gamma-irradiated cups with the most extensive abrasion increased in crystallinity, whereas the nonsterilized cups underwent a crystallinity decrease.
Medical subject headings
- Acetabulum
- Femur Head
- Polyethylenes
- Sterilization