Fractionated radiotherapy of small inoperable lesions of the brain using a non-invasive stereotactic frame.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 1869468.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Arteriovenous malformations (AVMs) and benign or low grade, small malignant tumors can be treated by stereotactic radiotherapy in a single fraction. This report describes a technique for stereotactic treatment of small lesions using conventional, fractionated, photon beam irradiation. The Laitinen's stereoadapter, non-invasive head frame was used. This device was tested for accuracy by serial mountings and found to be accurate within 1 mm. The accuracy of the dose delivered was within 2%. Adaptation of this device to the linear accelerator required the design of secondary circular collimators which decreased the penumbra from 3-4 mm to 2-3 mm. The dose fall off outside the target volume is steep enough when using two non-coplanar arcs (90 to 10% within 1 cm). Thermoluminescent dosimetry (TLD) in a humanoid phantom showed good correlation with the calculated dose. This system permits delivery of fractionated radiation therapy to small volumes, easily and accurately, under stereotactic conditions.
Medical subject headings
- Brain Neoplasms
- Intracranial Arteriovenous Malformations
- Radiotherapy Dosage
- Stereotaxic Techniques