Duration of nocturnal hypoglycemia before seizures.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 18694975.
- Also identified by DOI 10.2337/dc08-0863 and PMC identifier 2571056.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Despite a high incidence of nocturnal hypoglycemia documented by the use of continuous glucose monitoring (CGM), there are no reports in the literature of nocturnal hypoglycemic seizures while a patient is wearing a CGM device. In this article, we describe four such cases and assess the duration of nocturnal hypoglycemia before the seizure. In the cases where patients had a nocturnal hypoglycemic seizure while wearing a CGM device, sensor hypoglycemia (<60 mg/dl) was documented on the CGM record for 2.25-4 h before seizure activity. Even with a subcutaneous glucose lag of 18 min when compared with blood glucose measurements, glucose sensors have time to provide clinically meaningful alarms. Current nocturnal hypoglycemic alarms need to be improved, however, since patients can sleep through the current alarm systems.
Medical subject headings
- Blood Glucose
- Epilepsy
- Hypoglycemia