Eight- to ten-year clinical and radiographic outcome of a porous tantalum monoblock acetabular component.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 18703310.
- Also identified by DOI 10.1016/j.arth.2008.06.020.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In a prospective study, the authors used a porous tantalum monoblock acetabular component for primary total hip arthroplasty between November 1997 and June 1999. A total of 156 consecutive primary total hip arthroplasty were done in 143 patients younger than 75 years. A total of 151 hips had a follow-up time from 8 to 10 years. The average preoperative total Harris hip score of 44.0 +/- 13.8 increased to 97.0 +/- 6.2 at the latest follow-up. The average preoperative Oxford hip score of 43.3 +/- 6.5 improved to 13.9 +/- 2.3 at the latest follow-up. Radiographic evaluation including the Ein-Bild-Röntgen-Analyse (EBRA) digital system showed no radiographic evidence of gross polyethylene wear, progressive radiolucencies, osteolytic lesions, acetabular fracture, or component subsidence. There were 7 (4.5%) postoperative complications all unrelated to the acetabular component.
Medical subject headings
- Acetabulum
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Joint Diseases