Detection of one-lung intubation incidents.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 18720008.
- Also identified by DOI 10.1007/s10439-008-9550-8.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Lung sounds are very common source for monitoring and diagnosis of pulmonary function. This approach can be used for detecting one lung intubation (OLI) during anesthesia or intensive care. In this paper, an algorithm for detecting OLI from lung sounds is presented. The algorithm assumes a multiple-input-multiple-output system, in which a multi-dimensional auto-regressive model relates the input (lungs) and the output (recorded sounds). An OLI detector is developed based on the generalized likelihood ratio test (GLRT), assuming coherent distributed sources for each lung. This method exhibited reliable results also when the lungs were modeled by incoherent distributed sources, which is a more accurate model for lung sources. The algorithm was tested using real breathing sounds recorded in an operating room, and it achieved an OLI detection rate of more than 95%, for each breathing cycle.
Medical subject headings
- Algorithms
- Intubation
- Lung
- Respiratory Sounds