Motion artifact simulating aortic dissection on CT.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1872227.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We recently imaged two patients clinically suspected of having aortic dissection whose contrast-enhanced CT examinations, obtained on a new scanner with a 1-sec scanning time, showed findings suggesting an ascending aortic dissection. The subsequent clinical course and evaluation implied that the CT findings were predominantly artifactual. We identified identical artifacts in 18% of 50 consecutive contrast-enhanced CT examinations performed for a variety of indications on the same scanner. The double-lumen artifact, simulating an intimal flap, occurs in the proximal ascending aorta and is limited to one or two contiguous transaxial images. The artifact was not detected on two other CT units. We believe the artifact arises from motion of the aortic wall and the surrounding pericardial recesses during image acquisition.
Medical subject headings
- Aortic Dissection
- Aortic Aneurysm
- Aortography
- Tomography, X-Ray Computed