Transient penetration of the hip joint during in situ cannulated-screw fixation of slipped capital femoral epiphysis.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1874769.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Before the routine intraoperative use of fluoroscopy at our institution during procedures to stabilize a slipped capital femoral epiphysis, twenty-five patients (thirty hips) had in situ cannulated-screw stabilization of a slipped capital femoral epiphysis with use of biplane radiography. Thus, a permanent record of the procedure was available for review. Fourteen hips in fourteen patients had an intraoperative episode during which the joint was penetrated by the guide-pin assembly or cannulated screw, or both. This penetration was corrected at the time of the operation. Eleven patients were followed for a minimum of two years (mean, thirty-nine months; range, twenty-four to sixty-three months). All physes went on to closure. None of the patients had clinical or radiographic evidence of chondrolysis. Transient penetration of the hip joint did not lead to chondrolysis in this series. This suggests that a single episode of penetration by a pin or screw, with immediate removal from the joint, is not associated with the development of chondrolysis.
Medical subject headings
- Bone Screws
- Epiphyses, Slipped
- Hip Injuries
- Intraoperative Complications
- Wounds, Penetrating
Anatomy
- hip
- femur