In vivo wear rate of the Birmingham Hip Resurfacing arthroplasty. A review of 10 retrieved components.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 18757168.
- Also identified by DOI 10.1016/j.arth.2008.06.022.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The wear of 8 femoral and 2 acetabular components of the Birmingham Hip Resurfacing (Midland Medical Technologies Ltd, Birmingham, UK) was measured using a coordinate measuring machine (CMM5; SIP, Geneva, Switzerland). After a median in situ time of 13 months, measurements of the Birmingham Hip Resurfacing femoral heads revealed a median volumetric wear of 2.9 mm(3). In one case with a cup abduction angle of 70 degrees , a significantly higher wear of 17.8 mm(3) was found. The 2 acetabular components showed volumetric wear of 5.4 and 27.6 mm(3) after 14 and 15 months in vivo. The wear rates of the investigated femoral components were somewhat, but not significantly, higher than the previously investigated 28-mm Metasul heads (median, 0.8 mm(3); Zimmer GmbH, Winterthur, Switzerland).
Medical subject headings
- Arthroplasty, Replacement, Hip
- Equipment Failure Analysis
- Hip Prosthesis
- Prosthesis Failure