Mean-square error due to gradiometer field measuring devices.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 1879849.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Gradiometers use spatial common mode magnetic field rejection to reduce interference from distant sources. They also introduce distortion that can be severe, rendering experimental data difficult to interpret. Attempts to recover the measured magnetic field from the gradiometer output will be plagued by the nonexistence of a spatial function for deconvolution (except for first-order gradiometers), and by the high-pass nature of the spatial transform that emphasizes high spatial frequency noise. Goals of a design for a facility for measuring biomagnetic fields should be an effective shielded room and a field detector employing a first-order gradiometer.