Control of specific absorption rate distribution using capacitive electrodes and inductive aperture-type applicators: implications for radiofrequency hyperthermia.

Kato, H; Hand, J W; Prior, M V; Furukawa, M; Yamamoto, O; Ishida, T · IEEE Trans Biomed Eng · 1991

biomechanical · Level V

Where this comes from

Abstract

A method of controlling the specific absorption rate (SAR) distribution in radiofrequency hyperthermia is proposed. The superposition of the current density associated with capacitively coupled electrodes and that associated with H-field coupled inductive aperture-type applicators modifies the actual current distribution in the heating material. Using a two-dimensional finite element method, we have shown that "focusing" is possible such that the SAR at the center of a phantom can be adjusted to be approximately twice that in superficial regions, even though the wavelength is considerably greater than the dimensions of the phantom.

Medical subject headings