In vitro and finite element analysis of a novel rotator cuff fixation technique.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 18823800.
- Also identified by DOI 10.1016/j.jse.2008.06.002.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this investigation was to assess the biomechanic strength and stress dispersion at the repair site of surface-holding repair techniques for rotator cuff repair compared to the double-row technique. Eighteen bovine infraspinatus tendons were repaired using 3 different repair techniques: double-row repair, surface-holding repair with transosseous sutures, and surface-holding repair with knotless anchors. Biomechanical testing and two-dimensional finite element analysis were performed. The surface-holding repair with transosseous sutures provided 87.9% more stiffness than the double-row repair. The number of tendon-suture site failures of the surface-holding repair with transosseous sutures was smaller than the other 2 techniques. The finite element analysis showed that the surface-holding repair model had a more dispersing stress pattern compared to a double-row repair model. It suggests that these repair techniques can prevent high stress concentration compared to the double-row repair.
Medical subject headings
- Finite Element Analysis
- Rotator Cuff
- Rotator Cuff Injuries
- Suture Techniques
Anatomy
- shoulder