Penetration of a metallic femoral head through the acetabular shell.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 18848421.
- Also identified by DOI 10.1016/j.arth.2008.09.001.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Extensive wear of a metal-on-polyethylene total hip arthroplasty may rarely result in erosion of the metal-backed acetabular shell and penetration of the femoral head. We report on the case of an 85-year-old man who presented to the emergency department with an apparent dislocated total hip. He subsequently was discovered to have a periprosthetic fracture after an attempt at closed reduction of what was only discovered intraoperatively to be an irreducible transacetabular component central dislocation. Recognition of this rare complication may change clinical outcome.
Medical subject headings
- Acetabulum
- Hip Prosthesis
- Periprosthetic Fractures
- Prosthesis Failure
Anatomy
- femur
- hip
- pelvis