Three-dimensional analysis of computed tomography-based navigation system for total knee arthroplasty: the accuracy of computed tomography-based navigation system.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 18848426.
- Also identified by DOI 10.1016/j.arth.2008.07.007.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We evaluated the postoperative alignment of 37 primary total knee arthroplasties performed using a computed tomography-based navigation system (Vector Vision Knee 1.5; Brain Lab, Germany) with a new 3-dimensional analysis. The mean coronal femoral angle was 89.0 degrees +/- 1.4 degrees (85.5 degrees -92.8 degrees ), and the coronal tibial component was 89.2 degrees +/- 1.0 degrees (87.4 degrees -91.6 degrees ). The hip-knee-ankle angle was observed to be 178.2 degrees +/- 1.5 degrees (173.9 degrees -181.8 degrees ). The external rotational alignment of the femoral component relative to the surgical epicondylar axis was -0.5 degrees +/- 1.7 degrees (-3.2 degrees to 3.4 degrees ). The results demonstrated that a computed tomography-based navigation system provided a reasonably satisfactory component alignment. The discrepancy between the 2-dimensional and 3-dimensional evaluations was 1.0 degrees +/- 0.9 degrees (0.1 degrees -3.4 degrees ). Three-dimensional analysis is necessary to evaluate the accuracy of the navigation system.