Analysis of the mechanism of fracture of silicone metacarpophalangeal prostheses.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 18936133.
- Also identified by DOI 10.1177/1753193408093808.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Eleven fractured Sutter metacarpophalangeal prostheses were obtained from the hands of three patients. All of the implants had fractured at the junction of the distal stem and the hinge. After visual examination, the fracture faces were prepared and viewed using a scanning electron microscope. These images indicated that fracture was initiated by abrasion on the dorsal aspect of the distal stem of the prostheses, close to the hinge. Crack propagation was shown to be due to a fatigue process. Once a crack started, its direction of travel could be followed, using topographical features common to engineering fracture analyses. Propagation was from radial to ulnar and from dorsal to palmar.
Medical subject headings
- Equipment Failure Analysis
- Joint Prosthesis
- Metacarpophalangeal Joint
- Silicones