Interactive two-dimensional fresh blood imaging: a feasibility study.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 18982330.
- Also identified by DOI 10.1007/s00330-008-1218-0.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper presents work on interactive 2D projection magnetic resonance angiography, based on fresh blood imaging (FBI), which integrates navigation and triggering optimisation into a continuous fluoroscopic procedure. The technique was developed on a clinical 1.5-T MRI system and performed in five healthy volunteers. Initial feasibility study compared FBI projection angiograms with maximum intensity projection reformats from multi-slice, ECG-gated, 2D time-of-flight (TOF) in the lower peripheral arteries. A technical performance evaluation of 40 vessel segments showed that FBI generated angiograms of comparable diagnostic quality (P < 0.074) with fewer artefacts (P < 0.003). Quantitative vessel-to-background contrast measurements were higher in FBI (P < 0.014). The technique has potential application as an interactive vascular imaging tool in interventional or multi-location MRI examinations.
Medical subject headings
- Diagnostic Imaging
- Magnetic Resonance Angiography