A method for monitoring vascular access function during hemodialysis.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 19052534.
- Also identified by DOI 10.1038/ki.2008.581.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We tested a new bedside method to determine the function of native arteriovenous fistula in 16 patients performed during hemodialysis without stopping the treatment. We initially measured vascular access flow (Q(a)) in each patient using the Transonic HD01(plus) device. We then measured the pressure in arterial and venous drip chambers at different blood pump flow rates (Q(bset)=0, 50, 100, 250, 300, 350 ml/min). The intravascular blood pressure gradient (P(f)) between arterial and venous puncture sites was estimated by a mathematical model. P(f) was positive for low Q(bset), but became negative when Q(bset) overcame the threshold value (Q(Inv)). Such critical flow showed a high correlation with Q(a), even if it was systemically lower. Computer analysis of fluid dynamics showed that when the blood pump flow overcame the Q(Inv) threshold, a critical transition from laminar flow to vortex circulation took place downstream of the venous needle, causing a dangerous shearstress on the vessel wall. Our results show that Q(Inv) provides an indication of the maximal blood pump flow rate needed to be reached to maximize blood flow supply in order to limit hemodynamic stress on the vascular access.
Medical subject headings
- Arteriovenous Shunt, Surgical
- Evaluation Studies as Topic
- Renal Dialysis