Migration of the trabecular metal monoblock acetabular cup system.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 19056211.
- Also identified by DOI 10.1016/j.arth.2008.09.027.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Fifty-one primary total hip arthroplasties were performed using Trabecular Metal Monoblock Acetabular Cup System (Zimmer Inc, Warsaw, IN). In a 2-year prospective study, its behavior was closely monitored by clinical, radiologic, and component migration assessment by Ein-Bild-Röntgen-Analyse method. There were no complications. The mean Harris hip score was improved from 41 to 95. A polar gap at the postoperative radiograph was noted at 25% of the arthroplasties, most of which subsequently filled with bone within 6 months. The mean total absolute migration was 0.67 mm. The implant showed excellent early clinical and radiographic behavior. The 2-year migration rate study, as index of long-term survival and success, showed very good early implant stability and, in all cases except one, superior to the available studies for similar design acetabular cups.
Medical subject headings
- Acetabulum
- Arthroplasty, Replacement, Hip
- Foreign-Body Migration
- Hip Prosthesis
- Metals
- Prosthesis Failure
Anatomy
- hip
- pelvis