Minimum 5-year results in 96 consecutive hips treated with a tapered titanium stem system.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 19056225.
- Also identified by DOI 10.1016/j.arth.2008.09.020.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We performed total hip arthroplasty with single titanium stem in 96 consecutive, nonselected hips. All patients, regardless of bone type and shape, neck shaft angulation, or age, received the same implant. Patients were followed for a minimum of 5 years, and an independent radiographer evaluated the hips for ingrowth, subsidence, leg-length discrepancy, and remodeling. The average Harris Hip score was 96 points (range, 73-100 points) at final follow-up. Radiographically, all stems were ingrown. No stem had more than 3 mm of subsidence, and there were no leg-length discrepancies more than 5 mm. We concluded that this titanium stem is a versatile option for total hip arthroplasty in patients with a wide variety of demographic and femur characteristics.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Prosthesis Design
- Titanium
Anatomy
- hip