Tibial fixation without screws in cementless knee arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 19056226.
- Also identified by DOI 10.1016/j.arth.2008.09.023.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study reexamines the clinical and radiologic results of the first 216 cases of a cementless fixed-bearing tibial tray without screw holes in 191 patients at 7.6-year follow-up (range, 5-10 years). Mean Hospital for Special Surgery scores improved from 54/100 preoperatively to 84/100 postoperatively. Four tibial baseplates were revised. Five polyethylene inserts were revised. The overall reoperation rate was 12%. Four percent of tibial baseplates had incomplete nonprogressive radiolucent lines involving 1 or 2 of 10 defined zones. Overall tibial osteolysis rate was 4.1%. All tibial components were stable by radiographic technique. We conclude that this tibial tray provides reliable stability without the use of screws at medium-term follow-up. The polyethylene failure rate is a concern.