The outcomes of unicompartmental knee arthroplasties (UKAs) have demonstrated inconsistent long-term survival. We report the first clinical series of UKA using a semiactive robotic system for the implantation of an inlay unicondylar knee arthroplasty. Ten patients were selected for this study. Preoperative mechanical leg alignment values ranged from 0.3 degrees varus to 9.8 degrees varus. A haptic guidance system was used; a detailed description is given in the manuscript. The setup time for the robot was 41 minutes; intraoperative registration process, 7.5 minutes (6-13 minutes); skin incision, 8 cm; robot-assisted burring, 34.8 minutes (18-50 minutes); mean tourniquet time, 87.4 minutes (68-113 minutes); and overall operation time, 132 minutes (118-152 minutes). The planned and intraoperative tibiofemoral angle was within 1 degrees. The postoperative long leg axis radiographs were within 1.6 degrees. Haptic guidance in combination with a navigation module allows for precise planning and execution of both inlay components in UKA.