Confirmation of endotracheal tube placement: a miniaturized infrared qualitative CO2 detector.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 1905886.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A miniaturized, infrared, solid-state, end-tidal CO2 detector was used to confirm emergency endotracheal tube (ETT) placement. This prospective, clinical study used a miniature, infrared, solid-state end-tidal CO2 detector to confirm ETT placement in an acute setting. The ICU, emergency department, and hospital floor. There were 88 consecutive adult patients requiring 100 emergency intubations. The indication for airway intervention was considered urgent in 79% and under arrest conditions in 21%. The mean number of intubation attempts was 1.83 (range, one to five) with difficulty of intubation of 6.48 and confirmation of 7.75, on a linear scale from 0 (lowest) to 10 (highest). Determination of ETT position revealed intratracheal intubation in 96% and esophageal intubation in 4%. Placement was confirmed by direct visualization or radiography in all cases. Sensitivity and specificity for ETT localization was 100% (P less than .0001). This hand-held infrared capnometer reliably confirms ETT placement under emergency conditions.
Medical subject headings
- Carbon Dioxide
- Intubation, Intratracheal