Kinematics of a cementless mobile bearing posterior cruciate ligament-retaining total knee arthroplasty.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 19073365.
- Also identified by DOI 10.1016/j.knee.2008.11.008.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Mobile bearing (MB) total knee arthroplasty (TKA) was developed to provide low contact stress and unconstrained joint motion. We studied a consecutive series of 41 knees with mobile-bearing, posterior cruciate-retaining (CR) TKAs to determine if kinematics resembled normal knees or if kinematics changed over time. Patients were studied at 3 and 21 months average follow-up with weight-bearing radiographs at full extension, 30 degrees flexion and maximum flexion. Shape-matching techniques were used to measure TKA kinematics. Implant hyperextension, maximum flexion and total ROM increased with follow-up. Tibial rotation and condylar translations did not change with time. The medial condyle did not translate from extension to 30 degrees, but translated 5 mm anteriorly from 30 degrees to maximum flexion. Lateral condylar translation was 3 mm posterior from extension to 30 degrees, with no translation from 30 degrees to maximum flexion. Tibiofemoral kinematics in CR-MB-TKAs were stable over time, but did not replicate motions observed in healthy knees. The mobile tibial insert showed rotation and translation at both follow-up examinations, but the patterns of translation were not predictable.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Joint
Anatomy
- knee