Comparison of A-scan and MRI for the measurement of axial length in silicone oil-filled eyes.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 19074920.
- Also identified by DOI 10.1136/bjo.2008.147868.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The aim of the study was to compare the accuracy of A-scan biometry and MRI for the measurement of axial length in silicone oil-filled eyes. This was a prospective randomised study of 70 patients. Biometry was performed using MRI in 33 patients (MRI group) and A-scan echography in 37 patients (A-scan group). The difference between predicted and final refraction was measured and evaluated statistically. In patients with axial length >/=26 mm, the mean deviation of the final from predicted refraction was -1.23 (SD 0.67) D in the MRI group and -2.3 (SD 2.02) D in the A-scan group. The difference between these two groups was statistically significant (p = 0.02). In patients with axial length <26 mm, the mean deviation of the final from predicted refraction was -0.12 (SD 1.29) D in the MRI group and -0.33 (SD 1.39) D in the A-scan group. There was no statistical significance between the two groups (p = 0.629). For highly myopic patients MRI biometry was a more accurate measurement of axial length in silicone oil-filled eyes. A-scan and MRI biometry were comparably accurate in measuring axial length in patients with axial length <26 mm.
Medical subject headings
- Eye
- Refractive Errors
- Silicone Oils