Does increased bone-cement interface strength have negative consequences for bulk cement integrity? A finite element study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 19085106.
- Also identified by DOI 10.1007/s10439-008-9616-7.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Implant loosening is one of the most important modes of failure of cemented total hip replacement. It may be related to the cement strength, cement-prosthesis interface, cement-bone interface, surgical technique, or stem design. The main purpose of this study is to investigate the effect of bone-cement interface mechanical properties on cement degradation. The computational methodology proposed herein combines a previously developed bone-cement interface damage model and an accumulative damage model for bulk cement. This has been applied to a finite element model of an Exeter cemented hip implant. A higher strength of the bone-cement interface due to a higher amount of interdigitated bone results in faster cement deterioration. Over time, damage both at the bone-cement interface and in the cement mantle worsens. Also, a larger debonded area was predicted proximally, as observed in clinical practice. We conclude that the computational model proposed herein allows a realistic simulation of the bone-cement interface debonding and cement degradation, being a useful tool in the design of this kind of medical devices.