Orthogonal field calibration analysis for myocardial electrode arrays used in defibrillation studies.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19126463.
- Also identified by DOI 10.1109/TBME.2008.2003089.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Mapping the myocardial electric field during a defibrillation pulse requires the recording of potential differences between electrodes. The field is then calculated from these quantities and the corresponding calibration matrix. One straightforward calibration technique involves alignment of a known electric field along each of the orthogonal axes of an electrode array and recording the resulting potential differences. However, no results have been reported to support the efficacy of this technique. This study performs a detailed error analysis including a one-to-one comparison to a precision calibration technique, and quantitatively establishes the efficacy of the orthogonal field technique.