Novel MEA platform with PDMS microtunnels enables the detection of action potential propagation from isolated axons in culture.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19156289.
- Also identified by DOI 10.1039/b806689b and PMC identifier 2790813.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study investigated a novel multi-electrode-array (MEA) design capable of long-term and highly selective recordings of axonal signals using PDMS microtunnels. We successfully grew neurons in culture so that only axons extended through narrow (10 microm wide by 3 microm high) and long (750 microm) microtunnels under which multiple electrodes were integrated. This permitted the recording of relatively large (up to 200 microV) electrical signals, including the propagation speed and direction of these travelling action potentials. To further demonstrate the operation of the device as a diagnostic tool for drug screening assays, the drug mepivacaine was applied in washout experiments. Here, we identified significant changes in mean spiking rate and conduction velocity.
Medical subject headings
- Action Potentials
- Axons
- Dimethylpolysiloxanes
- Microtechnology
- Neural Conduction
- Tissue Array Analysis