Measurement of total blood flow in the normal human retina using Doppler Fourier-domain optical coherence tomography.
Where this comes from
- Record sourced from PubMed, PMID 19168468.
- Also identified by DOI 10.1136/bjo.2008.150276 and PMC identifier 2743389.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To measure total retinal blood flow in normal human eyes using Doppler Fourier-domain optical coherence tomography (FD-OCT). 10 normal people aged 35 to 69 years were measured for the right eye using Doppler FD-OCT. Double circular scans around the optic nerve heads were used. Four pairs of circular scans that transected all retinal branch vessels were completed in 2 s. Total retinal blood flow was obtained by summing the flows in the branch veins. Measurements from the eight scans were averaged. Veins with diameters >33 microm were taken into account. Total retinal blood flow could be measured in eight of 10 subjects: mean (SD) = 45.6 (3.8) microl/min (range 40.8 to 52.9 microl/min). The coefficient of variation for repeated measurements was 10.5%. Measured vein diameters ranged from 33.3 to 155.4 mum. The averaged flow speed was 19.3 (2.9) mm/s, which did not correlate with vessel diameter. There was no significant difference between flows in the superior and inferior retinal hemispheres. Double circular scanning using Doppler FD-OCT is a rapid and reproducible method to measure total retinal blood flow. These flow values are within the range previously established by laser Doppler flowmetry.
Medical subject headings
- Retinal Vessels