Prospective randomized trial comparing alumina ceramic-on-ceramic with ceramic-on-conventional polyethylene bearings in total hip arthroplasty.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 19195824.
- Also identified by DOI 10.1016/j.arth.2009.01.013.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This prospective randomized study aims to compare the outcome between an alumina ceramic-on-ceramic (CC) articulation with a ceramic on ultra-high-molecular-weight polyethylene articulation (CP). Fifty-six hips in 55 patients with mean age 42.2 (range, 19-56) each received uncemented components, a 28-mm alumina head with randomization of acetabular liner. Mean St Michael's outcome score for each group with up to 10 years follow-up (median, 8 years; range, 1-10) was 22.8 and 22.9, respectively (P = .819). Wear was identified in all but 1 CP replacement, but only 12 of the 23 CC. Mean wear in the CP group was 0.11 mm/y and 0.02 mm/yr in the CC group (P < .001). Other than significantly greater wear in the polyethylene group, there was no significant difference in midterm outcome between the 2 groups.
Medical subject headings
- Aluminum Oxide
- Arthroplasty, Replacement, Hip
- Bone Diseases, Developmental
- Ceramics
- Hip Prosthesis
- Osteoarthritis, Hip
- Polyethylene
Anatomy
- hip