Optimum receiver array design for magnetic induction tomography.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19203883.
- Also identified by DOI 10.1109/TBME.2009.2013936.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Magnetic induction tomography (MIT) is an imaging modality that aims at mapping the distribution of the electrical conductivity inside the body. Eddy currents are induced in the body by magnetic induction and the resulting fields are measured by an array of receiver coils. In MIT, the location of the receivers affects the quality of the image reconstruction. In this paper, a fast deterministic algorithm was applied to obtain optimum receiver array designs for a given specific excitation. The design strategy is based on the iterative exclusion of receiver locations, which yield poor conductivity information, from the space spanning all possible locations until a feasible design is reached. The applicability of "regionally focused" MIT designs that increase the image resolution at a particular region was demonstrated. Currently used design geometries and the corresponding reconstructed images were compared to the images obtained by optimized designs. The eigenvalue analysis of the Hessian matrix showed that the algorithm tends to maintain identical conductivity information content sensed by the receivers. Although the method does not guarantee finding the optimum design globally, the results demonstrate the practical usability of this algorithm in MIT experimental designs.
Medical subject headings
- Electromagnetic Fields
- Image Processing, Computer-Assisted
- Tomography