The effect on minimum tumor dose of restricting target-dose inhomogeneity in optimized three-dimensional treatment of lung cancer.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 1924862.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
An examination was made of the effect upon the minimum tumor dose of a limit placed on the variation of dose across target. If the required level of target dose uniformity is slightly relaxed, a substantial improvement in the minimum tumor dose might appear. It was conjectured that this effect could be seen with treatments optimally planned and evaluated in three-dimensions. A model of advanced carcinoma of the lung treated with a computer controlled accelerator was used to test this hypothesis. A mathematical program for optimizing beam weights was used to determine the largest minimum tumor dose possible. In the six cases tested, a minimum tumor dose of greater than 80 Gy could be delivered if a 20% inhomogeneity limit was accepted. The minimum tumor dose fell to the range 44-64 Gy when the inhomogeneity limit was tightened to 13-17%. The results imply a need to examine the choice of a required level of dose uniformity from the range of values suggested in the 2-dimensional planning literature. If a strict bound-on-dose uniformity is preserved, mechanisms--such as formal optimization--which can reduce target dose inhomogeneity will be valuable.
Medical subject headings
- Dose-Response Relationship, Radiation
- Lung Neoplasms
- Radiotherapy, Computer-Assisted