Fatal hemorrhage after attempted treatment of a basilar artery stenosis. Case report.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 19249928.
- Also identified by DOI 10.3171/2009.1.JNS08343.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Interventional treatment of severe intracranial stenoses using angioplasty and stent placement carries a significant risk of severe adverse events. These events are frequently caused by the intraarterial dilation but may also be precipitated by other causes related to the procedure. The authors describe a fatal subarachnoid hemorrhage during the treatment of a basilar artery (BA) stenosis caused by a vascular rupture distant from the site of angioplasty. After successful balloon dilation, an attempt to deploy a self-expanding stent failed due to the severe kinking of the stenotic segment of the BA. The angiogram performed during the intervention revealed a fatal rupture of the BA in a location not related to the angioplasty or guide wire tip. The technical aspects of the intervention are analyzed on the basis of the postmortem findings.
Medical subject headings
- Angioplasty
- Basilar Artery
- Cerebral Hemorrhage
- Stents
- Vertebrobasilar Insufficiency