Pressure-frequency sensing subxiphoid access system for use in percutaneous cardiac electrophysiology: prototype design and pilot study results.

Tucker-Schwartz, Jason M; Gillies, George T; Scanavacca, Mauricio; Sosa, Eduardo; Mahapatra, Srijoy · IEEE Trans Biomed Eng · 2009

Level II

Where this comes from

Abstract

We have designed, built, and tested an early prototype of a novel subxiphoid access system intended to facilitate epicardial electrophysiology, but with possible applications elsewhere in the body. The present version of the system consists of a commercially available insertion needle, a miniature pressure sensor and interconnect tubing, read-out electronics to monitor the pressures measured during the access procedure, and a host computer with user-interface software. The nominal resolution of the system is < 0.1 mmHg, and it has deviations from linearity of < 1%. During a pilot series of human clinical studies with this system, as well as in an auxiliary study done with an independent method, we observed that the pericardial space contained pressure-frequency components related to both the heart rate and respiratory rate, while the thorax contained components related only to the respiratory rate, a previously unobserved finding that could facilitate access to the pericardial space. We present and discuss the design principles, details of construction, and performance characteristics of this system.

Medical subject headings