Validation of continuous glucose monitoring in children and adolescents with cystic fibrosis: a prospective cohort study.

O'Riordan, Stephen M P; Hindmarsh, Peter; Hill, Nathan R; Matthews, David R; George, Sherly; Greally, Peter; Canny, Gerard; Slattery, Dubhfeasa et al. · Diabetes Care · 2009

prospective_cohort · Level II

Where this comes from

Abstract

To validate continuous glucose monitoring (CGM) in children and adolescents with cystic fibrosis. Paired oral glucose tolerance tests (OGTTs) and CGM monitoring was undertaken in 102 children and adolescents with cystic fibrosis (age 9.5-19.0 years) at baseline (CGM1) and after 12 months (CGM2). CGM validity was assessed by reliability, reproducibility, and repeatability. CGM was reliable with a Bland-Altman agreement between CGM and OGTT of 0.81 mmol/l (95% CI for bias +/- 2.90 mmol/l) and good correlation between the two (r = 0.74-0.9; P < 0.01). CGM was reproducible with no significant differences in the coefficient of variation of the CGM assessment between visits and repeatable with a mean difference between CGM1 and CGM2 of 0.09 mmol/l (95% CI for difference +/- 0.46 mmol/l) and a discriminant ratio of 13.0 and 15.1, respectively. In this cohort of children and adolescents with cystic fibrosis, CGM performed on two occasions over a 12-month period was reliable, reproducible, and repeatable.