Ceramic-ceramic bearing decreases osteolysis: a 20-year study versus ceramic-polyethylene on the contralateral hip.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 19283439.
- Also identified by DOI 10.1007/s11999-009-0773-2 and PMC identifier 2866910.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Although ceramic implants have been in use for many years and they are intended to minimize wear debris it is unknown whether alumina-on-alumina or alumina-on-polyethylene produce less wear and osteolysis. We therefore investigated wear and osteolysis on 28 bilateral arthroplasties (one ceramic-ceramic and the contralateral ceramic-polyethylene) of patients who had survived 20 years without revision and without loosening of either hip. Osteolysis was identified on anteroposterior pelvic radiographs and 3-D volume from CT scans. The number of osteolytic lesions detected with CT scan was higher than with radiographs. The number of lesions was higher on the side with the alumina-PE couple. With a similar length of followup on each side, the surface and the volume of osteolysis were consistently higher on the side with the alumina-PE couple. We found no correlation between the volume of osteolysis and the volume of estimated wear in each couple of friction. Hips with osteolysis had a lower Harris score.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Ceramics
- Hip Prosthesis
- Osteolysis
- Polyethylene
- Prosthesis Failure
Anatomy
- hip