The use of the Penlon Nuffield 200 in a monoplace hyperbaric oxygen chamber. An evaluation of its use and a clinical report in two patients requiring ventilation for carbon monoxide poisoning.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 1928680.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We modified a Penlon Nuffield 200 for use in a monoplace hyperbaric oxygen chamber by feeding back the chamber pressure to the reducing valve of the Nuffield 200. This provides a compensating mechanism, allowing the ventilator to deliver adequate tidal volumes at pressures of up to 3 atmospheres. We report the laboratory testing of the ventilator and our experience of ventilating two patients with carbon monoxide poisoning. Although compensation is not complete the modification is adequate for short-term clinical use in patients in whom the airway is compromised but who need hyperbaric oxygen therapy.
Medical subject headings
- Carbon Monoxide Poisoning
- Hyperbaric Oxygenation
- Ventilators, Mechanical