Analysis of bacterial detection in whole blood-derived platelets by quantitative glucose testing at a university medical center.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 19289590.
- Also identified by DOI 10.1309/AJCPVN9OT4GRCSKO.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
After the March 2004 implementation of American Association of Blood Banks standards regarding platelet bacterial detection, we began quantitative glucose screening of whole blood-derived platelets (WB-P). The glucose level was measured immediately before component release--often storage day 4 or 5--using the Glucometer SureStep Flexx Meter (LifeScan, Milpitas, CA), with a positive cutoff of less than 500 mg/dL; failing units were cultured and not transfused. During 29 months (March 1, 2004-July 31, 2006) 93,073 units of WB-P were tested. Initially, 929 units (0.998%) screened positively. Bacterial growth was culture-confirmed in 6 units, for a bacterial contamination incidence of 0.006% and a true-positive rate of 6.4/100,000. Three additional culture-confirmed contamination cases were detected in transfused units causing febrile nonhemolytic reactions, for a false-negative rate of 3.2/100,000. Our overall contamination prevalence was 9.6/100,000 units of platelets transfused, lower than ordinarily cited, and showed a false-negative rate remarkably congruent to that of culture: 3.2/100,000. A low-sensitivity screening test applied late in platelet shelf-life can be comparable to culture in preventing bacterial-related morbidity.
Medical subject headings
- Bacterial Infections
- Blood Platelets
- Glucose
- Platelet Transfusion
- Quality Assurance, Health Care