The effect of drill hole size on the fixation strength of a cemented prosthetic patellar button.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 19292359.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study used a cadaver model to analyze the fixation strength of 3 different patellar resurfacing preparations using an all-polyethylene patellar component. One of 3 drill hole sizes was randomly selected and used to prepare the patella. The entire construct was cemented, mounted to a servohydraulic testing machine, cycled between 0 and 50 N, and then tested to failure. Mean forces obtained prior to failure were 258.5, 293, and 353.1 N for the chamfer, 4.5-mm, and 9.5-mm drill holes, respectively. There was a statistically significant difference in strength to failure between the 9.5-mm and chamfer drill hole sizes. These findings may help to reduce patellar implant failures.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Cementation
- Osteotomy
- Patella
Anatomy
- knee