Accuracy of navigation: a comparative study of infrared optical and electromagnetic navigation.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 19298030.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We evaluated the accuracy of navigation systems for measuring the mechanical axis in patients undergoing total knee arthroplasty and in the synthetic bone model. Infrared optical and electromagnetic navigation systems were compared. Both systems were found to be accurate and reproducible in an experimental environment. However, the accuracies of both systems were affected by erroneous registration, and the optical system was found to be more reproducible. In clinical situations, the mean difference was 1.23 degrees, and difference greater than 3 degrees occurred in 15% of clinical trials. These discordances may have been due to ambiguous anatomic landmarks causing registration errors and the possibility of electromagnetic signal interference in the operating room.
Medical subject headings
- Diagnostic Imaging
- Electromagnetic Phenomena
- Femur
- Optical Devices
- Surgery, Computer-Assisted
Anatomy
- femur