Performance of four carbon dioxide absorbents in experimental and clinical settings.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19302642.
- Also identified by DOI 10.1111/j.1365-2044.2008.05759.x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the performance of four kinds of carbon dioxide (CO(2)) absorbents (Medisorb GE Healthcare, Amsorb Plus Armstrong Medical, YabashiLime Yabashi Industries, and Sodasorb LF Grace Performance Chemicals), we measured their dust production, acceptability of colour indicator, and CO(2) absorption capacity in in vitro experimental settings and the concentration of compound A in an inspired anaesthetic circuit during in vivo clinical practice. In vitro, the order of the dust amount was Sodasorb LF > Medisorb > Amsorb Plus = YabashiLime both before and after shaking. The order of the color acceptability was similar: Sodasorb LF > Amsorb Plus = Medisorb > YabashiLime both initially and 16 h after CO(2) exhaustion. During exposure to 200 ml.min(-1) CO(2) in vitro, the period until 1 kg of fresh soda lime allowed inspired CO(2) to increase to 0.7 kPa (as a mark of utilisation of the absorbent) was longer with Medisorb (1978 min) than with the other absorbents (1270-1375 min). In vivo, compound A (1.0% inspired sevoflurane) was detected only when using Medisorb. While Medisorb has the best ability to absorb CO(2), it alone produces compound A.
Medical subject headings
- Anesthesia, Closed-Circuit
- Carbon Dioxide
- Gas Scavengers