Allergy to pacemaker silicone compounds: recognition and surgical management.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 19324172.
- Also identified by DOI 10.1016/j.athoracsur.2008.08.028.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Silicone is a widely used biomaterial. Contact allergy, particularly to silicone components of pacemaker coatings, is uncommon. We present a 12-year-old girl with a history of complex congenital heart disease and acquired complete heart block excluding transvenous lead placement. Contact allergy to silicone led to multiple surgical interventions until the etiology for recurrent pacemaker wound complications was discovered. The key to diagnosis was a specific manufacturer's patch test. Complete removal of the former pacing system and placement of custom-made silicone free pacemaker components and epicardial use of silicone free transvenous leads were essential for successful therapy.
Medical subject headings
- Biocompatible Materials
- Dermatitis, Allergic Contact
- Pacemaker, Artificial