Malpositioned spinal instrument as a possible cause of superficial siderosis.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 19333243.
- Also identified by DOI 10.1038/sc.2009.35.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Case report. To report a patient with superficial siderosis as a complication after posterior fixation surgery for odontoid fracture. Department of Neurosurgery, Hokkaido University, Japan. A 36-year-old man had undergone C1-C2 posterior fixation using lamina hooks for an odontoid fracture in 1997. In 2003, he presented with hearing loss and ataxia; and in 2006, a diagnosis of superficial siderosis was made and spinal instrument malpositioning was detected. The malpositioned instrument, suspected as the cause of superficial siderosis, was removed. Superficial siderosis of the central nervous system is rare; it results in progressive hearing loss, cerebellar ataxia and pyramidal sign. Chronic hemorrhage in the subarachnoid space precipitates hemosiderin around the cerebellum and brainstem resulting in neurological symptoms. Recurrent hemorrhage and cervical root pathology, for example, root avulsion, are factors; the symptoms worsen gradually and result in hemostasis. Superficial siderosis because of complications from spinal instrumentation surgery is extremely rare. If the instrument is malpositioned in the subarachnoid space, we suggest its removal.
Medical subject headings
- Internal Fixators
- Postoperative Complications
- Siderosis
- Spinal Fractures
- Spinal Fusion
- Subarachnoid Hemorrhage