Secondary neutron doses for several beam configurations for proton therapy.

Shin, Dongho; Yoon, Myonggeun; Kwak, Jungwon; Shin, Jungwook; Lee, Se Byeong; Park, Sung Yong; Park, Soah; Kim, Dae Yong et al. · Int J Radiat Oncol Biol Phys · 2009

basic_science · Level V

Where this comes from

Abstract

To compare possible neutron doses produced in scanning and scattering modes, with the latter assessed using a newly built passive-scattering proton beam line. A 40 x 30.5 x 30-cm water phantom was irradiated with 230-MeV proton beams using a gantry angle of 270 degrees , a 10-cm-diameter snout, and a brass aperture with a diameter of 7 cm and a thickness of 6.5 cm. The secondary neutron doses during irradiation were measured at various points using CR-39 detectors, and these measurements were cross-checked using a neutron survey meter with a 22-cm range and a 5-cm spread-out Bragg peak. The maximum doses due to secondary neutrons produced by a scattering beam-delivery system were on the order of 0.152 mSv/Gy and 1.17 mSv/Gy at 50 cm from the beam isocenter in the longitudinal (0 degrees ) and perpendicular (90 degrees ) directions, respectively. The neutron dose equivalent to the proton absorbed dose, measured from 10 cm to 100 cm from the isocenter, ranged from 0.071 mSv/Gy to 1.96 mSv/Gy in the direction of the beam line (i.e., phi = 0 degrees ). The largest neutron dose, of 3.88 mSv/Gy, was observed at 135 degrees and 25 cm from the isocenter. Although the secondary neutron doses in proton therapy were higher when a scattering mode rather than a scanning mode was used, they did not exceed the scattered photon dose in typical photon treatments.