Flow-through functionalized PDMS microfluidic channels with dextran derivative for ELISAs.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 19370243.
- Also identified by DOI 10.1039/b816018j.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this work, a dextran modified PDMS microfluidic ELISA device was fabricated. The dextran functionalization was conducted with a simple, economic and fast flow-through process in a fabricated PDMS microfluidic device, and demonstrated significant enhancement of hydrophilicity and efficient covalent immobilization of proteins on the PDMS microchannel surface. The device was used to simultaneously detect multiple important biomarker IL-5, HBsAg, and IgG, showing a limit of detection of 100 pg mL(-1) and a dynamic range of 5 orders of magnitude, which significantly improved the performance of the reported hydrophobic and plasma-treated hydrophilic PDMS flow-through immunoassay devices. The fabricated PDMS device demonstrated its capability for colorimetric detection of proteins through direct observation by human eyes. Thus, this work not only demonstrates great potential to fabricate an economical and sensitive lab-on-chip system for high throughput screening of various infectious diseases, but also provides an opportunity to develop a portable microfluidic ELISA device via human eye examination for heath point-of-care services.
Medical subject headings
- Biosensing Techniques
- Dimethylpolysiloxanes
- Enzyme-Linked Immunosorbent Assay
- Flow Injection Analysis
- Microfluidic Analytical Techniques