Biomechanical evaluation of different designs of glenospheres in the SMR reverse total shoulder prosthesis: range of motion and risk of scapular notching.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 19393929.
- Also identified by DOI 10.1016/j.jse.2009.01.015.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Reverse total shoulder arthroplasty is a treatment option for cuff tear arthropathy. Scapular notching remains a concern. This biomechanical study compared the range-of-motion in different designs of glenospheres and hence the relative risk of scapular notching. A precision coordinate device was used to investigate four different designs of glenospheres (SMR prosthesis); 36 mm concentric (Standard), 36 mm eccentric, 44 mm concentric, and 44 mm eccentric glenospheres. The centre of rotation in each design was first established. The position of the humeral prosthesis was recorded in the plane of the scapula to compare the degree of adduction and the total range-of-motion. Eccentric glenospheres were found to improve range-of-motion by allowing a higher degree of adduction. Larger diameter glenospheres were found to improve range-of-motion by increasing adduction and abduction. Compared to the 36 mm concentric (standard) glenosphere, the 36 mm eccentric glenosphere improved adduction by 14.5 degrees, the 44 mm concentric glenosphere improved adduction by 11.6 degrees, the 44 mm eccentric glenosphere improved adduction by 17.7 degrees. Eccentric glenospheres with a center-of-rotation placed more inferiorly were shown to improve adduction. This design may reduce the clinical incidence of scapular notching.
Medical subject headings
- Arthroplasty, Replacement
- Biomechanical Phenomena
- Joint Prosthesis
- Prosthesis Design
- Scapula
- Shoulder Joint
Anatomy
- shoulder
- scapula