Review article: Validity of the KT-1000 knee ligament arthrometer.
review · Level V
Where this comes from
- Record sourced from PubMed, PMID 19398799.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The KT-1000 knee arthrometer (KT-1000) is an objective instrument to measure anterior tibial motion relative to the femur for anterior cruciate ligament (ACL) reconstruction. Four studies between 1950 and 2007 regarding validity of the KT-1000 were identified using a Medline search. One had interpretable information on sensitivities, specificities, and predictive values to validate the instrument as a diagnostic tool in patients with acute or chronic ACL injuries. Three had limitations in methodology. We suggest that the KT-1000 should be used with caution as an objective instrument. Rather, using a KT-1000 score derived by subtracting the anterior tibial motion relative to the femur of the injured knee to that of the uninjured knee may be more appropriate as a dichotomous diagnostic test with a threshold of 2 or 3 mm.